Filter and mask blockage tester
Face Masks and Respiratory Clogging Tester
Model: RF4318AB
The filter and mask obstruction tester is used to test the amount of dust collected on the filter when the airflow passes through the filter under a certain dust environment and the airflow passes through the filter to reach a certain breathing resistance.
The RF4318AB tester is composed of a dust generator, a dust test bin, a vacuum pump, and a flow meter. It is controlled by a single-chip computer and displayed on a touch screen.
Scope of application Half-mask respirators, filters, etc.
Applicable standards
| EN 143 | EN 149 |
Specifications:
- Cross section of dust test cabin: 650mmx650mm;
- Airflow of test cabin: 60m3/h, linear velocity 4cm/s;
- Dust concentration: 400±100mg/m3;
- Sampling rate of dust concentration: 2L/min;
- 2L tidal volume of breathing simulator (adjustable);
- Frequency of breathing simulator: 15 times/min (adjustable);
- The air flow of vacuum pump is 95L/min (adjustable);
- Dust medium: dolomite powder;
- Built-in humidifier and heater to control exhalation temperature and humidity;
- Exhalation temperature in the mask: 37±2℃, relative humidity 95%;
- Touch screen control, simple operation;
- Configure dust generator;
- Automatically control the dust concentration, flow rate and line speed;
- A variety of fixtures, suitable for respirator and mask testing;
- Built-in flowmeter;
- Automatic control of pressure and test time;
- Overpressure protection;
- Built-in 11 test standard test modes;
Order article number RF4409P Protective clothing synthetic blood penetration tester
RF4409P/1 Anti-synthetic blood penetration test head
RF4409P/2 block screen
RF4409P/3 cutting template (75x75mm)
RF4409P/4 torque wrench
RF4409P/5 stainless steel tray